PN: C193306T

COMPONENTS:

1x Ab-conjugated beads (premixed). PN: C193306TA. One bottle containing 1.5 mL of the following AimPlex antibody-bead conjugates:

S4P1 - anti-human Eotaxin

S4P4 - anti-human IP-10

S4P10 - anti-human IL-8

S5P1 - anti-human MCP-1

S5P3 - anti-human MCP-3

S5P7 - anti-human RANTES

2x Biotin-detection Ab (premixed). PN: C193306TB. One vial containing 0.45 mL of the following biotinylated antibodies:

Biotin - anti-human Eotaxin

Biotin - anti-human IP-10

Biotin - anti-human IL-8

Biotin - anti-human MCP-1

Biotin - anti-human MCP-3

Biotin - anti-human RANTES

2x NR dAb Diluent. PN: 07NDX21T. One vial containing 0.45 mL of biotin detection antibody diluent.

Lyophilized Standard Mix- Human Chemokine 6-Plex. PN: HG19306. One vial containing lyophilized recombinant Eotaxin, IL-8, IP-10, MCP-1, MCP-3, RANTES. 

      Standard Info Sheet: Indicates reconstitute volume and standard concentrations.

Basic Kit for Premixed Panels - 32 Tests: 

      1x Streptavidin-PE (SAPE). PN: 07SP0011T. One bottle containing 1 mL of 1x SAPE.

      10x Reading Buffer. PN: 07RB1T. One bottle containing 1.5 mL of 10x buffer.

      10x Wash Buffer. PN: 07WB1T. One vial containing 5 mL of 10x buffer.

      PCR 8-tube Strip. PN: 07P8T1. 2 strips.

      Filter Plate and Lid. PN: 07FPL1. 1 96- well filter plate and 1 plate lid.

      Plate Seals. PN: 07PSR1. 3 sealers.

STORAGE:  2-8 C in the dark.

IMPORTANT: Sodium azide forms explosive compounds with heavy metals. These products contain <0.05% (w/w) azide which with repeated contact with lead and copper commonly found in plumbing drains may result in the buildup of shock sensitive compounds. Dispose in accordance with regulations from your institute.

APPLICATION: Optimal antibody pairs and antigen standards for assaying assaying Human Chemokine 6-Plex Panel.  To be used in conjunction with the AimPlex Basic Kit for Premixed Panels - 32 Tests (PN: U100005T) and a diluent kit. Refer to the AimPlex Multiplex Immunoassay User Manual and kit inserts for the assay procedure

Assay Specifications:

Sample types: Cell culture supernatant, serum, plasma, bodily fluid and tissue/cell lysate

Sensitivity (LODs): Eotaxin, < 5 pg/mL; IL-8, < 1 pg/mL; IP-10, < 3 pg/mL; MCP-1, < 2 pg/mL; MCP-3, < 2 pg/mL; RANTES, < 5 pg/mL.

LLOQs: Eotaxin, < 10 pg/mL; IL-8, < 1 pg/mL; IP-10, < 10 pg/mL; MCP-1, < 5 pg/mL; MCP-3, < 5 pg/mL; RANTES, < 10 pg/mL.

ULOQs: Eotaxin, > 2,000 pg/mL; IL-8, > 1,000 pg/mL; IP-10, > 2,000 pg/mL; MCP-1, > 5,000 pg/mL; MCP-3,  > 5,000 pg/mL; RANTES,  > 5,000 pg/mL..

Standard dose recovery: 70-130%

Intra-assay CV: < 10%

Inter-assay CV: < 20%

Cross-reactivity of analytes in the Panel: Negligible

Sample volume: 15 µL/test